Z-0181-2021: Opticross 35 15 MHz Peripheral Imaging Catheter 8F (2.74 mm) X 105 cm Expiration Date June 8th 2021 - Jul

Recall record Z-0181-2021 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0181-2021
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Oct 22, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Opticross 35 Peripheral Imaging Catheter is being recalled because the incremental telescoping length markings near the proximal end of the catheter shaft can flake off when the device is manipulated. No patient harm has been reported to date. The most common potential adverse health consequence would be prolonged procedure to exchange the catheter for another as this potential issue is observable and has occurred outside of the patient's body. The most server health consequence that is reasonably expected to occur if a catheter with this potential issue is used is embolization which would result of the flaking marks advance while flushing the catheter. All recalled devices should be returned to Boston Scientific.

Source facts

Action
All affected customers were sent a letter on 08/24/2020 stating the following: Opticross 35 Peripheral Imaging Catheter is being recalled because the incremental telescoping length markings near the proximal end of the catheter shaft can flake off when the device is manipulated. No patient harm has been reported to date. The most common potential adverse health consequence would be prolonged procedure to exchange the catheter for another as this potential issue is observable and has occurred outside of the patient's body. The most server health consequence that is reasonably expected to occur if a catheter with this potential issue is used is embolization which would result of the flaking marks advance while flushing the catheter. All recalled devices should be returned to Boston Scientific.
Initiated
2020-08-24
Root cause
Process control
Terminated
2024-04-05
Distribution
US Nationwide distribution including in the states of (1) MD (2) NY (3) MI (4) CT (5) SD (6) GA.
Recall number
Z-0181-2021
Recall status
Terminated
CDRH record ID
183757
FDA record created
2020-10-22
Recall event number
86466

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗