Skip to content

Public FDA record

Z-0181-2022: Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151

Recall record Z-0181-2022 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0181-2022
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Oct 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c results may affect consideration of intervention.

Source facts

Action
On September 20, 2021, the firm distributed Urgent Medical Device Correction letters to impacted customers. Atellica customers were informed that the resolution of the issue will be implemented in SW v1.25.1. ADVIA customers were informed that the resolution of the issue will be implemented in AP Tool v1.11 and TDef v.1.07. These updates will be available soon. In the interim, customers were provided additional instructions to work around the issue until the update is available. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Initiated
2021-09-20
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide distribution.
Recall number
Z-0181-2022
Recall status
Open, Classified
CDRH record ID
189523
FDA record created
2021-10-28
Recall event number
88784

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗