The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.
Source facts
Action
On October 4, 2022, the firm notified affected customers via Important Medical Device Information letters. Customers were reassured that the product was manufactured according to specification and remains potassium-free. Customers may continue to use the product.
Please contact FMCRTG Customer Service at 1-800-323-5188 if you have further questions.
Initiated
2022-10-04
Root cause
Error in labeling
Distribution
Domestic distribution to AL
AZ
CA
DC
FL
GA
HI
IL
IN
MD
MI
MO
MS
NE
NJ
NY
OH
PA
SD
TX
VA
No foreign distribution.
Recall number
Z-0181-2023
Recall status
Open, Classified
CDRH record ID
196258
FDA classified
2022-11-04
Recall event number
90947
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.