Skip to content

Public FDA record

Z-0181-2023: NaturaLyte Liquid Acid concentrate, Part Number 08-0231-4, four (4) one-gallon bottles per bottle case.

Recall record Z-0181-2023 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0181-2023
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Nov 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The "Nominal Chemical Composition" section of the label incorrectly lists 5.62g/L KCL (potassium chloride). 5.62g/L actually corresponds to the CaCl (calcium chloride) content. The product remains potassium free.

Source facts

Action
On October 4, 2022, the firm notified affected customers via Important Medical Device Information letters. Customers were reassured that the product was manufactured according to specification and remains potassium-free. Customers may continue to use the product. Please contact FMCRTG Customer Service at 1-800-323-5188 if you have further questions.
Initiated
2022-10-04
Root cause
Error in labeling
Distribution
Domestic distribution to AL AZ CA DC FL GA HI IL IN MD MI MO MS NE NJ NY OH PA SD TX VA No foreign distribution.
Recall number
Z-0181-2023
Recall status
Open, Classified
CDRH record ID
196258
FDA classified
2022-11-04
Recall event number
90947

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗