Z-0181-2025: Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100

Recall record Z-0181-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0181-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Oct 23, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.

Source facts

Action
An URGENT: MEDICAL DEVICE CORRECTION notification letter dated 9/20/24 was sent to customers. Actions Required: Our records indicate that your facility has purchased one or more of the affected products. Therefore, Olympus requires you to take the following actions: 1. Carefully read the content of this notification. 2. Examine your inventory and identify the above-listed device with the affected serial number. 3. Beginning October 28, 2024, please call Olympus Customer Service at 1-800-848-9024, option 3, to request a CCD Color Adjustment . Olympus will arrange for the return of your device to our Olympus Repair Center at a mutually convenient time. Olympus Repair will perform a color adjustment. 4. If you have further distributed this product, identify your customers, and forward them this notification. 5. Olympus requests that you acknowledge receipt of this letter through the Olympus web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0454 c. Complete the form as instructed. Olympus requests that you report any complaints, including any imaging issues, to the Olympus Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1), and the FDA. Adverse events experienced with the use ofthis product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by email at Cynthia.Ow@Olympus.com.
Initiated
2024-08-21
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-0181-2025
Recall status
Open, Classified
CDRH record ID
210210
FDA classified
2024-10-23
Recall event number
95421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗