Z-0182-2007: MIDAS TOUCH Electrosurgical/102 mm (4 inch) Blade Electrodes, Disposable, Sterile Single Use Device, Part

Recall record Z-0182-2007 for Olsen Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0182-2007
Company in FDA record
Olsen Medical
FDA classification date
Nov 15, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Testing conducted by the firm indicates that electrodes identified with certain part numbers may fall out from the electrosurgical pencil or handle presenting the potential for harm or injury to the patient and/or user.

Source facts

Action
The firm issued an ''Urgent: Device Recall'' notification to their customers via letter and E-mail on 8/29/2006 and 8/30/2006. The letter informs the customers of this problem and asks that the customers contact their sub-accounts and notify them of the problem if they have further distributed any of the suspect part numbers. The letter also asks that the distributor/customer(s) have the subaccounts discontinue use of the product; and return the held product to the distributor (who in turn, should return the product to Olsen Medical for reimbursement). The customers are asked to complete and fax back to the recalling firm, an attached Acknowledgement form documenting receipt of the recall notification by the customer/employee with name and date received. The firm also asked that the customer supply information regarding the amount of inventory /shipped on a checklist. The amounts received are broken down on the Checklist by Lot Number with the amount of product in inventory and the amount of product shipped. These amounts are reported by the responders by ''pieces and cases'' on hand at the time that the notice was received.
Initiated
2006-08-29
Root cause
Other
Terminated
2011-07-26
Distribution
Worldwide, including USA, Japan, and New Zealand.
Recall number
Z-0182-2007
Recall status
Terminated
CDRH record ID
48702
FDA classified
2006-11-15
Recall event number
36470

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗