Circlip component used to suspend flat panel and navigation arm system may become dislodged.
Source facts
Action
Notification letter was sent to consignees September 5, 2008. Field service will be performed at all affected sites. Old Circlips will be replaced with new clamps that do not allow the arm to fall. Contact Stryker Communications Corp. at 1-972-410-7310 for assistance.
Initiated
2008-09-05
Root cause
Device Design
Terminated
2009-10-02
Distribution
Nationwide and International Distribution
Recall number
Z-0182-2009
Recall status
Terminated
CDRH record ID
73708
FDA classified
2008-10-22
Recall event number
49514