Z-0182-2022: Oncentra Brachy users using version 4.0 and above with Applicator Modeling license 170283-00

Recall record Z-0182-2022 for Elekta, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0182-2022
Company in FDA record
Elekta, Inc.
FDA record creation date
Oct 28, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain circumstances, there can be a difference between the planned and delivered dose of the two ovoid channels.

Source facts

Action
An Urgent Important Field Safety Notification letter (FAC-NU-0007) was issued to consignees to notify of the issue, clinical impact and recommend the following: 1) Recommended User Action in the unique scenario, which is only applicable for the Venezia" applicator: Avoid use of the combination of the two Venezia" applicator Anchor Points O1-Center and O2-Center by disabling one of them in the Applicator Library Manager. 2)Recommended User Action in the exceptional situation that may also occur for other Applicator Models: Avoid incorrect placement of the Anchor Points, especially exchanging left / right anchor point positions. 3) Recommended User Actions for both scenarios: Until an updated version of Oncentra Brachy is available, it is strongly advised: " To use additional Anchor Points during applicator modelling. This will further reduce the chance of mirroring the Applicator Model. " To cross-check the source dwell position information from the Case Explorer to be in accordance with the source dwell positions in the image views. " Use the Triangles option for the applicator skin visualization in the 3D view, and review whether the Applicator Model has a correct shading display. " Perform proper Quality Assurance for all treatment plans before treatment delivery. Consignees are required to acknowledge receipt of the notification to Elekta within 30 days.
Initiated
2021-10-13
Root cause
Software design
Distribution
Distribution throughout USA OUS distribution including:Algeria, Argentina, Australia, Austria Bahamas, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil Brunei, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran (Islamic Republic of), Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Libyan Arab Jamahiriya, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam
Recall number
Z-0182-2022
Recall status
Open, Classified
CDRH record ID
189744
FDA record created
2021-10-28
Recall event number
88860

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗