Z-0182-2023: Sedecal Soltus Battery Mobile X-Ray System REF Soltus 450 Battery Mobile X-Ray System REF Soltus 500

Recall record Z-0182-2023 for SEDECAL SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0182-2023
Company in FDA record
SEDECAL SA
FDA classification date
Nov 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to un-commanded movement of mobile x-ray unit.

Source facts

Action
Sedecal SA provided Canon Medical Systems USA (CMSU) the sole distributor of the Soltus system in the U.S. an "URGENT: ADVISORY NOTICE" to communicated to CMSU's customers. On 07/28/2022, CMSU sent the "URGENT: ADVISORY NOTICE" Letter via USPS Certified to inform customer of an issue with erratic movements related with Soltus 500 Mobile X-Ray units. According to reports, the users were trying to turn the systems to the left or to the right when it suddenly started moving backwards at a high speed. Customer are instructed to: 1. Do not continue using the unit when in error condition (magenta LED status lights blinking) if the error cannot be reset. 2. In case of undesired or erratic movements perceived by the user, release the handle and park the unit immediately. Do not try to fight against the erratic movements of the system with the handle. 3. Call service immediately to restore the unit to operational status. 4. If the unit most be moved, power off the unit (do not use the emergency stop button) and use either the manual driving brake release button or the manual clutch screws (See Manual (OM-0474R11) sections 3.6.1.3 and 3.6.1.6 for model Soltus 500. See Manual (OM-0475R7) sections 3.6.1.3 and 3.6.1.5 for model Soltus 450. For questions, contact Gary Becker, Regulatory Affairs Manager at 800-421-1968 or gmbecker@us.medical.canon or the Canon Service Rep at 800-521-1968.
Initiated
2022-08-12
Root cause
Software design
Distribution
U.S.: KY, WI, FL, DE, WV, CA, MO, NY, NC, AL, AR, MS, GA, IL, IA, TN, KS, KY, IN, HI, OH, WY, PA, NV, RI, UT, AZ, OK, MN, ME, MT, WA, VA, SD, MI, LA, and TX O.U.S.: N/A
Recall number
Z-0182-2023
Recall status
Open, Classified
CDRH record ID
195606
FDA classified
2022-11-04
Recall event number
90736

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗