Open original FDA record ↗Recall record
- FDA identifier
- Z-0183-2008
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Jan 23, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Crack, Fracture or Detachment: The Transmitter Arm for the Mayfield-compatible Transmitter may become detached from the unit and enter into the surgical field.
Source facts
- Action
- Consignees were sent an Urgent Recall Notice via mail dated 10/11/07, notifying the customer to examine the plastic end of the transmitter arm and to not use the transmitter arm, if any cracks or damage are observed. Also to stop using Steam Sterilization method and to utilize the other Sterilization methods indicated in the Operators Manual. The root cause is still under investigation GEHC Surgery will provide, at no charge, a solution to this issue when it becomes available.
- Initiated
- 2007-10-18
- Root cause
- Device Design
- Terminated
- 2012-08-10
- Distribution
- Worldwide Distribution --- USA including states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IL, MA, MD, MI, MS, NC, NY, OK, OR, PA, SC, TX, UT, VA, WI, and WV, and countries of Canada, China, France, Hong Kong, Japan, Mexico, Spain, and Switzerland.
- Recall number
- Z-0183-2008
- Recall status
- Terminated
- CDRH record ID
- 65470
- FDA classified
- 2008-01-23
- Recall event number
- 45441