Open original FDA record ↗Recall record
- FDA identifier
- Z-0183-2015
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA classification date
- Nov 7, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A cable inside the C-arm systems may have been routed in an inappropriate manner. Improper routing of the cable may result in increased wear over time. Without additional measures, the potential exists for a cable inside the C-arm to break which could result in restricted functionality or system failure and an ongoing procedure could be terminated. Additionally, a subgroup of affected C-arm systems will receive a replacement of the cable harness.
Source facts
- Action
- Siemens sent an " Important Safety Information " letter dated September 23, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Siemens will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. Our service organization will contact with you to schedule an appointment to perform the corrective action. Please feel free to contact our service organization for an earlier appointment. Please observe this safety notice and comply with the corresponding measure until the update has been fully completed. For further questions, please call (610) 219-6300.
- Initiated
- 2014-09-23
- Root cause
- Device Design
- Terminated
- 2017-05-08
- Distribution
- Nationwide Distribution including Puerto Rico.
- Recall number
- Z-0183-2015
- Recall status
- Terminated
- CDRH record ID
- 130747
- FDA classified
- 2014-11-07
- Recall event number
- 69480