Z-0183-2023: Zio AT Clinical Reference Manual, Part: ALB0031, and Zio AT Important Information Pamphlet, Part: ALB0034

Recall record Z-0183-2023 for iRhythm Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0183-2023
Company in FDA record
iRhythm Technologies, Inc.
FDA classification date
Nov 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ambulatory ECG monitoring system labeling update: 1) Clinically actionable arrhythmia notifications delayed until patient registration is complete; 2) Patches limited to 100 symptomatic and 500 asymptomatic wireless transmissions; once the maximum transmission limit is reached for either type, any further transmissions for that type will cease. Risk of delayed reporting to health providers.

Source facts

Action
On 9/28/2022, Field Safety Notice Labeling Corrections were sent to customers who were informed of labeling changes and that the firm intends to enhance the process of notifying patients that they are approaching the maximum limit for either symptomatic or asymptomatic transmissions by auto shipping another Zio AT patch to the patient. Customers are asked to do the following: 1) Please review this letter and disseminate it to the appropriate system providers and any personnel at your facility. Please share this notification with all device users within your facility to ensure they are aware of this notice. 2) Complete and return the customer reply form. Further questions can be directed to customer care at (888) 693-2401 or support@irhythmtech.com or your local sales representative.
Initiated
2022-09-28
Root cause
Device Design
Distribution
US: MD, NY, FL, TN, WA, AL, MA, AZ, CA, KS, HI, PA, NJ, TX, MN, VA, AR, OR, NC, CO, KY, MS, LA, MI, WI, WV, IN, ID, SC, CT, SD, IA, NV, OK, OH, IL, VT, NH, NE, NM, MO, ME, DC, RI, MT, DE, GA, ND
Recall number
Z-0183-2023
Recall status
Open, Classified
CDRH record ID
196267
FDA classified
2022-11-04
Recall event number
90954

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.