Open original FDA record ↗Recall record
- FDA identifier
- Z-0183-2024
- Company in FDA record
- Carl Zeiss Suzhou Co., Ltd.
- FDA classification date
- Oct 26, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.
Source facts
- Action
- On 8/18/23, correction notices were mailed to customers who were asked to do the following: Before each use of the affected device, perform the following mandatory inspection: Refer to pictures in the notice showing a gap between the OPMI and the inspection arm. This gap should not be greater than 5mm. If the gap is greater than 5mm, please Immediately stop using the device! In this case the affected device can no longer be used. The firm is planning to inspect (repair if needed) the suspension arm immediately. Service staff will contact you to arrange an appointment for inspection (repair if needed) of the suspension arm. Complete and return the acknowledgement and receipt form and return to USFCA.meditec.us@zeiss.com If you have any questions, contact firm Customer Support at 877-486-7473.
- Initiated
- 2023-08-18
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of MD, TX, CA, GA, LA.
- Recall number
- Z-0183-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203536
- FDA classified
- 2023-10-26
- Recall event number
- 93146