Z-0183-2024: OPMI LUMERA 300, REF 6137

Recall record Z-0183-2024 for Carl Zeiss Suzhou Co., Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0183-2024
Company in FDA record
Carl Zeiss Suzhou Co., Ltd.
FDA classification date
Oct 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.

Source facts

Action
On 8/18/23, correction notices were mailed to customers who were asked to do the following: Before each use of the affected device, perform the following mandatory inspection: Refer to pictures in the notice showing a gap between the OPMI and the inspection arm. This gap should not be greater than 5mm. If the gap is greater than 5mm, please Immediately stop using the device! In this case the affected device can no longer be used. The firm is planning to inspect (repair if needed) the suspension arm immediately. Service staff will contact you to arrange an appointment for inspection (repair if needed) of the suspension arm. Complete and return the acknowledgement and receipt form and return to USFCA.meditec.us@zeiss.com If you have any questions, contact firm Customer Support at 877-486-7473.
Initiated
2023-08-18
Root cause
Process control
Distribution
US Nationwide distribution in the states of MD, TX, CA, GA, LA.
Recall number
Z-0183-2024
Recall status
Open, Classified
CDRH record ID
203536
FDA classified
2023-10-26
Recall event number
93146

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗