Skip to content

Public FDA record

Z-0183-2027: Convenience Kits containing CONMED AirSeal Access Ports CHN GYN ONCOLOGY DYNJ906731K DYNJ90673

Recall record Z-0183-2027 for Medline Industries, LP. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0183-2027
Company in FDA record
Medline Industries, LP
FDA classification date
Oct 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.

Source facts

Action
The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
Initiated
2026-08-31
Root cause
Process change control
Distribution
US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
Recall number
Z-0183-2027
Recall status
Open, Classified
CDRH record ID
222824
FDA classified
2026-10-15
Recall event number
99888

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗