Z-0184-2007: M1020B PulseOximetry Module Philips FAST Sp02 and Pulse Oximetry Nellcor OxiMax Compatible used with the

Recall record Z-0184-2007 for Philips Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0184-2007
Company in FDA record
Philips Medical Systems
FDA classification date
Nov 16, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Unexpected pulse oximetry (Sp02) readings (100%) over time when sensor is not attached to patient. In addition, with a sensor attached to a patient, an incorrect high Sp02 may be displayed when a pulse rate is in a range of 185BPM

Source facts

Action
Philips issued a Customer Letter and Urgent Device Correction letter to consignees on 09/21/06. Users were notified of the problem and provided with Instructions to Mitigate Risk. A software upgrade to: version D.00.61 and Sp02 FW A.01.46 is mandatory.
Initiated
2006-09-21
Root cause
Other
Terminated
2020-08-11
Distribution
Worldwide, including USA, Canada, Australia, China, Hong Kong, India, Indonesia, Japan, Phillipines, Singapore, Taiwan, Thailand, Vietnam, Argentina, Brazil, Chile, Colombia, El Salvador, Jamaica, Mexico, Nicaragua. Panama, Paraguay, Peru, Venezuela, Austria, Belgium, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hungary, Germany, Israel, Italy, Jordan, Kenya, Kuwait, Lebanon, Lithuania, Luxembourg, Macedonia, Mauritania, Mauritius, Netherlands, Norway, Oman, Slovenia, South Africa, Spain, Switzerland, Turkey, UAE, UK, Yemen, Zambia, Zimbabwe
Recall number
Z-0184-2007
Recall status
Terminated
CDRH record ID
48751
FDA classified
2006-11-16
Recall event number
36489

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗