Skip to content

Public FDA record

Z-0184-2014: 2008K@ Home Blood lines Combiset True Flow Series for Hemodialysis use Catalog Number: 03-2962-3. Intend

Recall record Z-0184-2014 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0184-2014
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Nov 8, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Updated Instructions for Use: Bloodline connections not tightened per instructions in the operator's manual may disconnect and lead to blood loss

Source facts

Action
Fresenius issued a Field Notification Letter dated 8/29/13 to Patient Clinics/ associated at home patients and provides additional detail instructions to ensure the connection to the transducer protector bloodline female luer interface is secure. This field advisory notice sent by Certified Mail, Return Receipt Requested. Accounts requested to return both the receipt and the completed reply form (enclosed) to confirm receipt of the notice. If you have any questions regarding this field advisory, please contact FMCNA Customer Service Team at 1-800-323-5188. Customer Service will be available Monday-Friday 7am-6pm CST.
Initiated
2013-08-29
Root cause
Labeling design
Terminated
2021-02-11
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Recall number
Z-0184-2014
Recall status
Terminated
CDRH record ID
120010
FDA classified
2013-11-08
Recall event number
65739

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗