Reason stated in the source

During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infection.

Source facts

Action
On 09/10/21 recall notices were mailed to customers. Action to be taken by the user 1. Immediately quarantine and discontinue use of the affected products. 2. Pass on this Urgent Field Safety Notice to all users of the products listed above and all other persons who need to be aware within your organization. 3. If you have distributed the affected products to third parties, please promptly forward this letter to the recipients of the products and indicate contact details of the recipient on the Acknowledgement and Response Form. 4. Return the Acknowledgment and Response Form by Fax or E-Mail to the contact listed below within 10 calendar days of receipt of this notice. 5. Get in touch with your KARL STORZ representative to return affected products and to discuss suitable alternative products. This notice must be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Complete and return the Customer Acknowledgement and Response Form.
Initiated
2021-09-10
Root cause
Under Investigation by firm
Distribution
US: SC, NH, UT, LA, FL, TX, MA, MO, VA, IA, CA, IL, NC, VT, AL, GA, PA, TN, WI, NY, CO, AZ, NV, ME, MS, DC, IN, KY, MD, OH, MI, WA, KS, OR, AR, MN, NJ, NM, WV, CT. OUS: Canada
Recall number
Z-0184-2022
Recall status
Open, Classified
CDRH record ID
189315
FDA record created
2021-10-29
Recall event number
88694

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗