Open original FDA record ↗Recall record
- FDA identifier
- Z-0184-2022
- Company in FDA record
- Karl Storz Endoscopy
- FDA record creation date
- Oct 29, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infection.
Source facts
- Action
- On 09/10/21 recall notices were mailed to customers. Action to be taken by the user 1. Immediately quarantine and discontinue use of the affected products. 2. Pass on this Urgent Field Safety Notice to all users of the products listed above and all other persons who need to be aware within your organization. 3. If you have distributed the affected products to third parties, please promptly forward this letter to the recipients of the products and indicate contact details of the recipient on the Acknowledgement and Response Form. 4. Return the Acknowledgment and Response Form by Fax or E-Mail to the contact listed below within 10 calendar days of receipt of this notice. 5. Get in touch with your KARL STORZ representative to return affected products and to discuss suitable alternative products. This notice must be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Complete and return the Customer Acknowledgement and Response Form.
- Initiated
- 2021-09-10
- Root cause
- Under Investigation by firm
- Distribution
- US: SC, NH, UT, LA, FL, TX, MA, MO, VA, IA, CA, IL, NC, VT, AL, GA, PA, TN, WI, NY, CO, AZ, NV, ME, MS, DC, IN, KY, MD, OH, MI, WA, KS, OR, AR, MN, NJ, NM, WV, CT. OUS: Canada
- Recall number
- Z-0184-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189315
- FDA record created
- 2021-10-29
- Recall event number
- 88694