Z-0184-2023: FERNO POWER X1 AMBULANCE COT- Item #0015807/UDI-DI 00190790002183 with 36V Li-Ion Batteries Model Number

Recall record Z-0184-2023 for Ferno-Washington Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0184-2023
Company in FDA record
Ferno-Washington Inc
FDA classification date
Nov 4, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Batteries could become unstable when not properly maintained and/or subjected to repeated drops, power washing directly into battery contacts or abuse, may result in a fire and/or injury

Source facts

Action
Ferno issued Urgent Product Correction on 10/7/22 via Certified Mail. Letter states reason for recall, health risk and action to take: Ferno will replace affected batteries with new batteries. Immediate Action. 1. Immediately check the serial number of your Ferno POWER X1 Ambulance Cot (see list within this notice) as well as the serial number on the bottom of any ENDURACharge 36V Li-Ion battery utilized with the cot. See How to Identify ENDURACharge Battery Date Code . 2. Check the serial number on the bottom of your ENDURACharge 36V Li-Ion battery to see if the battery is in the identified date code range. If it is not, no further action is required and the battery may remain in service provided it does not have a compromised seal, damaged case, cracks, deformities, other visual damage or is hot to the touch. Regardless of date code, if you identify a battery that has a compromised seal, damaged case, cracks, deformities, other visual damage or is hot to the touch, the battery should be immediately removed from service and moved outside your building. 3. ALL ENDURACHARGE 36V LI-ION BATTERIES MANUFACTURED WITHIN THE ABOVE REFERENCED DATE CODE RANGE MUST BE REPLACED. If you have a battery that is in the identified date code range, immediately visually inspect the battery. If you identify a battery that has a compromised seal, damaged case, cracks, deformities, other visual damage or is hot to the touch, the battery should be immediately removed from service and moved outside your building. As stated in the battery manual, A damaged battery can cause fire and/or injury. Inspect the battery before each use. If damaged, remove from service, recycle per the regulations in your country, and replace. If you have a battery that is in the identified date code range and it does not display any wear or damage and is operating as intended, the following actions should be immediately taken if use is continued while awaiting a replacement battery: " Do not leave the battery
Initiated
2022-10-07
Root cause
Component design/selection
Distribution
Nationwide Foreign: BRAZIL BRUNEI CHILE CHINA PERU POLAND PUERTO RICO TURKS&CACOS
Recall number
Z-0184-2023
Recall status
Open, Classified
CDRH record ID
196279
FDA classified
2022-11-04
Recall event number
90957

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗