Z-0185-2012: The LIFEPAK 20 defibrillator/monitor is an AC powered external defibrillator with battery backup used by

Recall record Z-0185-2012 for Physio Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0185-2012
Company in FDA record
Physio Control, Inc.
FDA classification date
Nov 16, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

When users do not disconnect the test plug and reconnect the QUIK-COMBO electrodes to the therapy cable, the test plug signal can be misinterpreted as a patient waveform.

Source facts

Action
Physio Control sent a " Urgent Medial Device Safety Alert" letter dated January 2009, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. The letter instructed the customers to follow the Operating Instructions for AED and Manual Defibrillation mode (Section 4-Therapy, pages 58 and 61) that instructs users to "Connect therapy electrodes to the therapy cable and confirm cable connection to the defibrillator" to prevent users from mistakenly leaving the test plug and reconnect the QUIK-COMBO electrodes to the therapy cable, the test plug signal can be misinterpreted as a patient waveform. Recommendations to users who use a Therapy Cable: -Disconnect the QUIK-COMBO test plug prior to patient use -Confirm therapy electrodes - to - therapy cable connection -Confirm therapy cable - to - defibrillator connection -Confirm therapy electrodes - to - patient connection Adding 2005 American Heart Association software that includes audible and visual prompts to LIFEPAK 20 defibrillators manufactured prior to January 2007 is optional and not required. If the customers would like to have this optional kit installed in your defibrillators, please contact Physio Control, Inc. Customers can call Technical Support at 1-800-442-1142 for any questions about this recall or or visit our website at www.physio-control notices.com/LP20testplug.
Initiated
2008-12-30
Root cause
Employee error
Terminated
2011-11-21
Distribution
Worldwide Distribution--USA ( nationwide ) and the countries of Bahrain, Lebanon , Cyprus, Netherlands, Egypt, Pakistan, Germany, Saudi Arabia, India, South Africa, Iran, Syria, Israel, Turkey, Jordan, UAE,Kuwait, Yemen, Asia Pacific, Australia, New Caledonia, Guam, New Zealand, Hong Kong, Latin America, Anguilla, Jamaica, Argentina, Mexico, Bahamas, Nicaragua, Brazil, Panama, Chile, Peru, Colombia, Trinidad, Costa Rica, Uruguay, El Salvador, Venezuela, Guatemala, Virgin Islands and Canada.
Recall number
Z-0185-2012
Recall status
Terminated
CDRH record ID
104736
FDA classified
2011-11-16
Recall event number
60202

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗