Skip to content

Public FDA record

Z-0186-2027: IntelePACS, InteleConnect Portal/Tech Portal Versions PACS-5-1-1-R102, PACS-5-2-1-R93, PACS-5-3-1-R83 thr

Recall record Z-0186-2027 for INTELERAD MEDICAL SYSTEMS INCORPORATED. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0186-2027
Company in FDA record
INTELERAD MEDICAL SYSTEMS INCORPORATED
FDA classification date
Oct 15, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Following a Chromium-based web-browser, report and displayed images may not match.

Source facts

Action
Intelerad, a GE HealthCare company, issued an urgent medical device correction notice to its consignees on 9/5/2026 via email. This notice contained a link error and was corrected and reissued on 9/9/2026. The notice explained the issue, potential risk and requested the following: "Actions to be taken by Customer/User Pending correction by Intelerad, you can continue to use IntelePACS - InteleConnect Portal and Technologist Portal by following the instructions below: 1. Do not use Google Chrome 152 or Microsoft Edge 152 browsers to access the Physician Portal Case Viewer: a. Use Firefox or Safari as the browser. b. Use an earlier version of the Chrome/Edge browser (versions preceding Google Chrome 152 and Microsoft Edge 152) 2. If you cannot use Firefox or Safari or an earlier version of the Chrome/Edge browser, always refresh the browser web page after opening each exam/study to ensure the correct report is displayed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via email or print, fill out manually, scan, and email to Recall.88102@gehealthcare.com." For questions or concerns regarding this notification, please contact Intelerad at Recall.88102@gehealthcare.com.
Initiated
2026-09-05
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-0186-2027
Recall status
Open, Classified
CDRH record ID
222843
FDA classified
2026-10-15
Recall event number
99899

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗