Z-0187-06: Guidant INSIGNIA I Entra family of pacemakers includes the following: SSI (model numbers 0484 and 0485);

Recall record Z-0187-06 for Guidant Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0187-06
Company in FDA record
Guidant Corporation
FDA classification date
Sep 24, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A failure mode occurring at a low rate, has been identified within the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include: Intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation.

Source facts

Action
Important Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05.
Initiated
2005-09-22
Root cause
Other
Terminated
2006-12-27
Distribution
Nationwide and Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Polynesia, Portugal, Qatar, Reunion, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom and Venezuela
Recall number
Z-0187-06
Recall status
Terminated
CDRH record ID
41956
FDA classified
2005-09-24
Recall event number
33491

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗