Open original FDA record ↗Recall record
- FDA identifier
- Z-0187-2008
- Company in FDA record
- Allez Spine, LLC
- FDA classification date
- Mar 11, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Screw Shank Failure; separation of the screw shank portion from the pedicle head portion of the Size 8 Pedicle Screw, when torquing down the single piece locking nut during implantation.
Source facts
- Action
- On July 30, 2007 Allez Spine notified all its field representatives and distributors via email of a potential for failure of the Size 8 Pedicle Screws when used in combination with the single piece locking nuts and instructed them to return all unused inventory of Size 8 Screws and the two piece Locking Nuts with saddle. This action was to be carried out to the retail level-hospital, clinic, lab, and physicians sold the product. The email also had an attachment of a list of expected frequently asked questions (FAQ) from surgeons and answers to help them respond. A second Notice to all users was sent by the recalling firm on Feb 29, 2008, to request follow-up of implanted units.
- Initiated
- 2007-07-30
- Root cause
- Device Design
- Terminated
- 2012-04-09
- Distribution
- Nationwide: including the states of AZ, CA, CO, GA, NV, TX and UT
- Recall number
- Z-0187-2008
- Recall status
- Terminated
- CDRH record ID
- 64520
- FDA classified
- 2008-03-11
- Recall event number
- 44911