Open original FDA record ↗Recall record
- FDA identifier
- Z-0187-2018
- Company in FDA record
- Physio-Control, Inc.
- FDA classification date
- Dec 12, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The infant artwork on the product instructs incorrect pad placement. Also, the artwork is missing the Type BF Patient connection and High Voltage Warning symbols.
Source facts
- Action
- The firm Physio Control, initiated their field action on 10/27/2017 by issuing a press release. The firm followed this with a recall notice dated 10/23/2017, to the direct consignees on 11/03/2017. The letter described the product, explained the problem and provided the following instructions: Quarantine the affected lots;contact Physio-Control at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday to arrange for replacement of your unused electrodes and "Upon receipt of this notification, post a copy of the enclosed Correct Electrode and Packaging Labeling with each of your AEDs, which shows the correct placement of the electrodes." Physio_Control will provide replacement products for all unused affected Electrodes. Should you have any questions about this Product Recall, please contact us at 1-866-231-1220, 6:00 A.M. to 4:00 P.M. (Pacific), Monday Friday.
- Initiated
- 2017-10-27
- Root cause
- Process control
- Terminated
- 2019-06-12
- Distribution
- Worldwide distribution: US (nationwide) including Puerto Rico and Guam and countries of: Canada, Austria, Belgium, Czech Republic; Denmark, Finland, France, French Polynesia, Germany, Hungary, Ireland, Italy, Latvia, Netherlands, Australia, China, Hong Kong, Japan, Argentina, Mexico, Norway, Panama, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom, and New Zealand.
- Recall number
- Z-0187-2018
- Recall status
- Terminated
- CDRH record ID
- 159097
- FDA classified
- 2017-12-12
- Recall event number
- 78303