Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Source facts
Action
Smiths medical issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 09/24/2024 via traceable mailing. The notice explained the issue, potential risk, and requested the destruction of all affected product. Distributors were directed to forward the notice to their customers.
Initiated
2024-09-24
Root cause
Process control
Distribution
Worldwide distribution.
Recall number
Z-0187-2025
Recall status
Open, Classified
CDRH record ID
210273
FDA classified
2024-10-23
Recall event number
95438