Open original FDA record ↗Recall record
- FDA identifier
- Z-0187-2026
- Company in FDA record
- American Contract Systems Inc.
- FDA classification date
- Oct 16, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Source facts
- Action
- Firm notified affected distributors who received affected product via email on 9/4/25. An updated letter was provided on 9/9/25, which better clarified product codes and impacted lot numbers and bag serial numbers. Consignees were instructed to immediately discontinue use and discard impacted trays. Distribution partners were provided with a customer letter for further distribution to end users.
- Initiated
- 2025-09-04
- Root cause
- Nonconforming Material/Component
- Distribution
- US Nationwide distribution in the states of SD, IA, MN, WA, IL.
- Recall number
- Z-0187-2026
- Recall status
- Open, Classified
- CDRH record ID
- 216324
- FDA classified
- 2025-10-16
- Recall event number
- 97659