Open original FDA record ↗Recall record
- FDA identifier
- Z-0188-2009
- Company in FDA record
- MAQUET Inc.
- FDA classification date
- Nov 5, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Table tilts: The Alphastar Surgical Table piston rod head may break and cause a full inclination of the table top during Trendelenberg positioning if there is insufficient lubrication.
Source facts
- Action
- Maquet sent Recall Notification letters to all customers on September 16, 2008 by Federal Express. Letters inform customers of the potential problem with the Alphastar tables, a description of the failure and hazards involved. Hospitals are asked to confirm whether yearly maintenance by an authorized technician have been performed. If preventive maintenance has not been performed, customers are requested to contact their Maquet Service rep. A Customer Service Response form is included and customers are to complete if to acknowledge whether of not the facility is still in possession of an affected table. If so, the model number, serial number and facility name & address is to be completed so that Maquet reps can visit the facility and perform the corrective action.
- Initiated
- 2008-09-16
- Root cause
- Equipment maintenance
- Terminated
- 2010-01-12
- Distribution
- Nationwide
- Recall number
- Z-0188-2009
- Recall status
- Terminated
- CDRH record ID
- 73516
- FDA classified
- 2008-11-05
- Recall event number
- 49448