Z-0188-2022: 3M Red Dot Electrodes, Radiolucent Monitoring Electrode with Foam Tape, Catalog No. 2244, (3M ID (SKU): 7

Recall record Z-0188-2022 for 3M Company - Health Care Business. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0188-2022
Company in FDA record
3M Company - Health Care Business
FDA record creation date
Nov 1, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for breakage of the ECG electrode carbon stud from the eyelet.

Source facts

Action
The recalling firm, 3M Company, issued "Urgent - Medical Device Recall" letters dated 29 September 2021 to the distributors on 10/6/2021. The letter informs them of the issue and instructs them to immediately discontinue distribution of the affected lots and determine if any are in inventory. The affected product is to be destroyed by the distributor. The distributor is to notify any customers who have purchased the affected lots by using the enclosed Customer Letter and Customer Acknowledgment Form provided. The enclosed customer letter to be sent to the distributor's customers is dated 29 September 2021 and informs the customer of the issue. The customer is instructed to screen their facility or other storage locations to ensure all affected lots of the product are identified. They are to immediately cease use of affected lots of product and quarantine it for destruction/disposal. The customer is requested to pass the notification along to all who need to be aware within their organization or to any department where the affected product was transferred.Complete the attached Customer Acknowledgement Form with the following: o Acknowledge that you have read and understood this letter, whether or not you have any affected lots of the 3M" Red Dot" Radiolucent Monitoring Electrodes with Foam Tape, Catalog Number 2244 in stock at your facility; and o Indicate if you have any product in stock and dispose of this material per facility policy. o Email the completed form to 3MFieldSafetyAction@montage72.com. " Complete and return the Customer Acknowledgement Form to 3M even if your stock of affected lots of this product have been exhausted. " To request product replacement for affected lots requiring disposal, contact 3M Health Care Customer Helpline at 1-800-228-3957, option #7 or by email to 3MFieldSafetyAction@montage72.com. If you have questions regarding this matter, please contact your local 3M sales representative or the 3M Health Care Customer Hel
Initiated
2021-10-06
Root cause
Process control
Terminated
2024-01-17
Distribution
US Nationwide Distribution to states of: AL, AZ, CA, IA, IL, LA, MD, MO, NC, NJ, and OH; and OUS Foreign distribution to countries of: Australia, Bermuda, Canada, China, France, Germany, Japan, the Netherlands, Singapore, Sweden, and Turkey.
Recall number
Z-0188-2022
Recall status
Terminated
CDRH record ID
189471
FDA record created
2021-11-01
Recall event number
88763

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗