Skip to content

Public FDA record

Z-0188-2024: StealthStation S8 Application version 2.0 and 2.0.1 (Part 9735762) The StealthStation System, with Steal

Recall record Z-0188-2024 for Medtronic Navigation, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0188-2024
Company in FDA record
Medtronic Navigation, Inc.
FDA classification date
Nov 10, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to a software anomaly which potentially could result in the surgical planning data shifting to an unintended location.

Source facts

Action
On 09/27/2023, the firm sent an "URGENT: MEDICAL DEVICE CORRETION" Letter/Communication via UPS to customer informing them of a software anomaly affecting certain StealthStation S8 and StealthStation FlexENT system with the StealthStation S8 App software version 2.0 and 2.0.1. Due to the software anomaly, there is the potential for the surgical planning data to encounter a displayed shift which could result in proceeding to an incorrect location from the planned target. This result in unintended tissue damage leading to permanent neurological injury, additional pass of device (biopsy needle or electrode), prolonged procedure or need for additional surgery. Customer are instructed to: 1. Follow the below instructions and refer to Appendix A for full details on impacted systems, issues and mitigations: a. Do not change the reference exam in Cranial or ENT procedures, if the surgical planning data is defined on a reference exam merged with pre-merge type or diffusion series. 2. Please review this information, including the additional details in Appendix A, with all physician users. If you have any questions related to this issue, please contact Medtronic Technical Services for help at 1-888-826-5603 or email at rs.navtechsupport@medtronic.com 3. Please confirm via the enclosed confirmation form that this notification has been communicated within your facility with all physician users. Send the completed Customer Confirmation Form to Medtronic via email at neuro.quality@medtronic.com. a. Medtronic will contact your facility to coordinate the software update when it is available. 4. This notice needs to be passed on to those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Questions or assistance, contact Medtronic Sales Representatives or Technical Services at 1-888-826-5603.
Initiated
2023-09-27
Root cause
Software design
Distribution
Worldwide Distribution: U.S. (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.(foreign) including countries of: ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BRAZIL, BULGARIA, CANADA, CANARY ISLANDS, CHILE, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, FRENCH POLYNES, GERMANY, GREECE, HONG KONG, HUNGARY, ICELAND, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KAZAKHSTAN, KENYA, KOREA, KUWAIT, LATVIA, LITHUANIA, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PANAMA, PHILIPPINES, POLAND, PORTUGAL, REUNION, ROMANIA, RUSSIAN FEDERATION, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UKRAINE, and UNITED KINGDOM
Recall number
Z-0188-2024
Recall status
Open, Classified
CDRH record ID
203577
FDA classified
2023-11-10
Recall event number
93158

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗