Z-0188-2026: Brand Name: MEGA Ballast Distal Access Platform Product Name: Mega Ballast Model/Catalog Number: MEGABT09

Recall record Z-0188-2026 for BALT USA, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0188-2026
Company in FDA record
BALT USA, LLC
FDA classification date
Oct 24, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Due to an identified manufacturing defect involving coating material lifting within the catheter's distal inner diameter/lumen.

Source facts

Action
On 09/18/2025, the firm sent via Priority Mail an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that, Balt USA has became aware of a situation where multiple Mega Ballast units failed the in-house QC final inspection due to the exterior coating material lifting. Further assessment found this coating also went inside of the ID and protruded from the inner surface of the catheter, which is not intended as part of the manufacturing process. Hospitals are instructed to: - Cease use of the affected product. - Inform hospital staff, such as safety officers, pharmacists, purchasing agent, head of neuroradiology and the neuroradiology staff, physicians including but not limited to endovascular neurosurgeons and interventional neurologists, as well as any other person if deemed necessary. - Identify and locate the Mega Ballast products concerned by this recall procedure. - Collect and put in quarantine the Mega Ballast products concerned by this recall, then return them to Balt USA through the usual administrative returns procedure by contacting Balt USA s Customer Service department. Distributors are instructed to: - Cease processing of the affected recall product. - Inform customers about this notice. - Identify and locate the Mega Ballast products concerned by this recall procedure. - Collect and put in quarantine the Mega Ballast products concerned by this recall procedure and then return them to your local distributor as per their return procedure. For questions or further assistance: Contact: Quality Department Email: QA@baltgroup.com Balt USA, LLC 29 Parker Suite 100 | Irvine CA, 92618 | USA Telephone: +949.788.1443 | Fax: +949.788.1444
Initiated
2025-09-25
Root cause
Process control
Distribution
US distribution to states of: AL, CA, CO, CT, FL, IL, KY, MA, NY, TN, TX, and UT.
Recall number
Z-0188-2026
Recall status
Open, Classified
CDRH record ID
216258
FDA classified
2025-10-24
Recall event number
97661

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗