Z-0189-2019: Atellica Sample Handler Prime (SMN 11069001). The Atellica" Solution is a multi-component system for in

Recall record Z-0189-2019 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0189-2019
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA record creation date
Oct 18, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacks and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.

Source facts

Action
Siemens issued an Urgent medical Device Correction, (UMDC ASW18-02.A.US) and Urgent Field Safety Notification (UFSN ASW18-02.A.OUS) on 6/29/18, via FedEx to the US Customer and via email distribution to countries Outside US (OUS). The field action letters explain the behaviors and informed Customers of action to be taken to prevent the behaviors from occurring. Siemens has released an updated software (SW) version (V), 1.14, which corrects the reported behaviors.
Initiated
2018-06-29
Root cause
Software design
Terminated
2021-08-12
Distribution
Nationwide Foreign: Australia Austria Belgium Brazil Canada Croatia Czech Republic Denmark France Germany Greece Italy Japan Malaysia Mexico Netherlands Norway Poland Republic Korea Singapore Spain Sweden Switzerland U.A.E. United Kingdom
Recall number
Z-0189-2019
Recall status
Terminated
CDRH record ID
166371
FDA record created
2018-10-18
Recall event number
80774

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗