Diagnostic ultrasound system with the specified
software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display
the measurement results correctly with two cardiology measurement functions:
Mitral Regurgitation Flow (MR Flow) Measurement;
Mitral Valve (MV) Measurement.
In addition there is an error in the instruction manual.
Source facts
Action
Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21.
Letter states reason for recall, health risk and action to take:
Customers/Users are advised to accept the corrected software and manual as soon as available.
Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems.
We will contact you to set up time to update the software and provide the revised manual. (7) Other Information:
For questions, comments, or concerns, please contact:
Aaron Pierce
Director, QA/RA
1959 Summit Commerce Park
Twinsburg, OH 44087
Phone (330) 425-1313 x2934, Fax (330) 963-0749
Email aaron.pierce@fujifilm.com
Initiated
2021-10-14
Root cause
Software design
Distribution
Nationwide
Recall number
Z-0189-2022
Recall status
Open, Classified
CDRH record ID
189704
FDA record created
2021-11-02
Recall event number
88840