Skip to content

Public FDA record

Z-0191-2020: Irrisept Wound Debridement and Cleansing System. 12 units per case.

Recall record Z-0191-2020 for IrriMAX Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0191-2020
Company in FDA record
IrriMAX Corporation
FDA record creation date
Oct 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential defect in the sterile pouch seal, which may result in a lack of sterility assurance.

Source facts

Action
Irrimax Corporation notified customers on about 08/16/2019 via "URGENT: MEDICAL DEVICE RECALL" letter. The letter informed customers that the recall was due to the potential of a defect in the sterile pouch seal, which may result in a lack of sterility assurance, and that the seal defect is not easily visible upon inspection of the pouch. Instructions to customers included, that within 10 business days of receiving the recall notification, they are to discontinue use of the recalled product (labeled lot numbers 19CAB939 or 19DAB431), complete and return the attached Recall Return Response Customer Acknowledgement and Receipt for to Irrimax by email (customercare@irrimax.com) or mail (Irrimax Corporation, 1665 Lakes Parkway, Suite 102, Lawrenceville, GA 30043), and arrange to return any unused affected product by following the Product Return Instructions provided. Distributors were instructed to discontinue use of the recalled product (labeled lot numbers 19CAB939 or 19DAB431), quarantine any inventory from the recalled lots to ensure product will no longer be distributed, complete and return the attached Recall Return Response Customer Acknowledgement and Receipt for to Irrimax by email (customercare@irrimax.com) or mail (Irrimax Corporation, 1665 Lakes Parkway, Suite 102, Lawrenceville, GA 30043), and provide a customer distribution list with customer, name, address, telephone numer and email address (contact information) and send to Irrimax via email (customercare@irrimax.com) by August 23, 2019. Irrimax will send the customer a letter and follow-up with the customer to return all unused product. Additionally, distributors were asked to return recalled product in inventory by following the Product Return Instructions provided. Questions can be directed to customercare@irrimax.com by email or toll free 1-844-477-4778.
Initiated
2019-08-16
Root cause
Process control
Terminated
2020-09-25
Distribution
Nationwide distribution to AL, AZ, CA, CO, FL, GA, HI, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV.
Recall number
Z-0191-2020
Recall status
Terminated
CDRH record ID
175870
FDA record created
2019-10-24
Recall event number
83599

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗