Open original FDA record ↗Recall record
- FDA identifier
- Z-0192-2018
- Company in FDA record
- HILL-ROM MANUFACTURING, INC.
- FDA record creation date
- Nov 30, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Monarch Electromagnetic Interference (EMI) may disrupt the operation of implanted pacemakers or implantable cardioverter defibrillators as well as other devices that might be subject to an EMI if they come into direct contact with the Monarch device.
Source facts
- Action
- Phone calls were initiated September 15, 2017, to the customers, who are users. The expanded recall was initiated by letter dated November 13, 2017,via FedEx with delivery confirmation to both the users identified in the September notification and to the users who later received the recalled product. (Two local consignees will be provided the letter in person and one consignee will be sent the letter via USPS.) For further questions, please call (800) 426-4224.
- Initiated
- 2017-09-15
- Root cause
- Other
- Terminated
- 2019-05-29
- Distribution
- USA (nationwide) Distribution to the states of: AZ, CA, CO, CT, FL, GA, IA, IL, IN, KY, MA, MI, MN, MO, NE, OH, OK, OR, TN, TX, VA, WA, WI, and WV. Internationally to Canada. There was no military/government distribution.
- Recall number
- Z-0192-2018
- Recall status
- Terminated
- CDRH record ID
- 158666
- FDA record created
- 2017-11-30
- Recall event number
- 78100