Z-0192-2022: Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0

Recall record Z-0192-2022 for FujiFilm Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0192-2022
Company in FDA record
FujiFilm Healthcare Americas Corporation
FDA record creation date
Nov 2, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual

Source facts

Action
Fujifilm issued Urgent Medical Device Correction letter dated 10/7/21, mailed 10/14/21. Letter states reason for recall, health risk and action to take: Customers/Users are advised to accept the corrected software and manual as soon as available. The Advisory Notification applies to all customers and distributors of the ARIETTA 65 system with software version V1.0 through V4.0.0. We will contact you to set up time to update the software and provide the revised manual. For questions, comments, or concerns, please contact: Aaron Pierce, Director, QA/RA 1959 Summit Commerce Park Twinsburg, OH 44087 Phone (330) 425-1313 x2934, Fax (330) 963-0749. Email aaron.pierce@fujifilm.com
Initiated
2021-10-14
Root cause
Software design
Distribution
US Nationwide distribution.
Recall number
Z-0192-2022
Recall status
Open, Classified
CDRH record ID
189722
FDA record created
2021-11-02
Recall event number
88849

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗