Z-0192-2024: EVIS EXERA III BRONCHOVIDEOSCOPE Model Numbers BF-1TH190, BF-H190, BF-Q190 and BRONCHOVIDEOSCOPE OLYMPUS

Recall record Z-0192-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0192-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Nov 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is a risk of endobronchial combustion if high-frequency cauterization is performed while supplying oxygen [and/or] the electrode section of the electrosurgical accessory is too close to the distal end of the endoscope.

Source facts

Action
An "URGENT MEDICAL DEVICE CORRECTIVE ACTION" notification letter dated 10/12/23 was sent to customers. Actions to be taken by the end user: Our records indicate that your facility has purchased one or more of the affected bronchoscopes. Olympus requests you to take the following actions: 1. Inspect your facility for the referenced devices and ensure all personnel are completely knowledgeable and thoroughly aware of the Warnings in affected bronchoscope s Operation Manual for use with High-frequency therapy equipment and that Olympus high-frequency compatible bronchoscopes are compatible only with Combination equipment list in operation manual. You may continue to use the device according to the existing instructions and warnings contained in the Instruction for Use. 2. Complete the enclosed response form and return to our recall partner, Sedgwick, via Email (Olympus4907@sedgwick.com) or Fax (866-808-1177). For any questions about the acknowledgement form, please call the Sedgwick team at 855-215-5128. 3. If you have further distributed this product outside of your facility, forward them this notification and appropriately document your notification process. Olympus requests that you report to Olympus complaints, including any injuries associated with procedures involving energy devices used with Olympus bronchoscopes. Please report complaints to the Olympus Technical Assistance Center (TAC) at 1-800-848-9024 (option 1). Adverse events experienced with the use of this product may also be reported to the Food and Drug Administration MedWatch Adverse Event Reporting program either online, by regular mail or by fax. Olympus regrets any inconvenience caused and fully appreciates your cooperation in this matter. Please do not hesitate to contact me directly at (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Initiated
2023-10-12
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution
Recall number
Z-0192-2024
Recall status
Open, Classified
CDRH record ID
203338
FDA classified
2023-11-09
Recall event number
93052

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗