Open original FDA record ↗Recall record
- FDA identifier
- Z-0193-06
- Company in FDA record
- Guidant Corporation
- FDA classification date
- Nov 23, 2005
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Another failure mode, which occurs at a low rate, has been identified with the INSIGNIA and NEXUS families of implantable pacemakers. Device behaviors which may be observed include; intermittent or permanent loss of pacing output without warning, intermittent or permanent loss of telemetry and appearance of a reset warning message upon interrogation.
Source facts
- Action
- Important Medical Device Safety Information & Corrective Action , dated 09/22/05 was sent to all affected physicians. This letter explained the failure mode, what devices are affected and gave recommendations. A press release was issued on 09/22/05. UPDATE: Guidant is issuing an Advisory Update to physicians in December 2005. This update provides information relative to the failure mode including the identified cause; FDA's classification; Guidant representatives will retrieve and replace any hospital inventory with product free from "Mode 2" peri-implant failure; and recommended actions.
- Initiated
- 2005-09-22
- Root cause
- Other
- Terminated
- 2007-01-05
- Distribution
- Nationwide and Arab Emerites, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Guadeloupe, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, Netherlands Antilles, New Zealand, Norway, Panama, Peru, Philippines, Polynesia, Portugal, Qatar, Reunion, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Swaziland, Sweden, Switzerland, Syrian Arab Republic, Taiwan ROC, Thailand, Turkey, United Kingdom, Urguay and Venezuela
- Recall number
- Z-0193-06
- Recall status
- Terminated
- CDRH record ID
- 42690
- FDA classified
- 2005-11-23
- Recall event number
- 33871