The ball screw on the unit which supports the detector head (which weighs approximately 800 lbs.) could snap resulting in the detector head dropping on the patient.
Source facts
Action
The recalling firm modified one of the devices in 2002. The remaining device users were telephoned on/about 11/02/2006.
To date the recalling firm has not taken any corrective action regarding devices with serial numbers B-102, B-103, B-104 or B-106 thru B-184.
Initiated
2006-08-01
Root cause
Other
Terminated
2008-10-29
Distribution
Worldwide, including USA, Australia, Egypt, UK, Belgium.
Recall number
Z-0193-2007
Recall status
Terminated
CDRH record ID
48531
FDA classified
2006-11-16
Recall event number
36381