Skip to content

Public FDA record

Z-0193-2013: Boule Con Diff US Tri Pack, 6 Vials. For in vitro diagnostic use as a control to monitor the performan

Recall record Z-0193-2013 for Clinical Diagnostic Solutions. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0193-2013
Company in FDA record
Clinical Diagnostic Solutions
FDA classification date
Nov 5, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class III

Reason stated in the source

Clinical Diagnostic Solutions, Inc. in Plantation, FL recalled the Boule Con Diff US Tri Pack (Product part 501-605, Lot #1005-549 and Product part 501-607, Lot #1005-549) which is used for in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments. The recall was imitated due to receipt of customers stating the MCV values and RDW% recovered for Boule

Source facts

Action
Clinical Diagnostic Solutions, Inc. sent a "PRODUCT NOTIFICATION:BOULE CON DIFF TRI-LEVEL LOT 1005-549" letter dated August 4, 2010 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. For questions regarding this notice, contact CDS Customer Service at 1-800-453-3328.
Initiated
2010-08-04
Root cause
Process control
Terminated
2014-08-14
Distribution
Nationwide Distribution.
Recall number
Z-0193-2013
Recall status
Terminated
CDRH record ID
112532
FDA classified
2012-11-05
Recall event number
63094

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗