Skip to content

Public FDA record

Z-0193-2018: pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blist

Recall record Z-0193-2018 for Pro-Med Instruments Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0193-2018
Company in FDA record
Pro-Med Instruments Gmbh
FDA record creation date
Dec 5, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip.

Source facts

Action
Pro-Med Instruments Gmbh initiated their recall by email on October 30, 2017. The recall notice stated the following: " Actions to be taken by the customer I user: Review this notification and ensure that all users of the affected products are informed of this urgent field safety notice. If you have transferred the affected products to third parties, please forward a copy of this letter or inform the below mentioned contact person. Customers were instructed to check their stock for any affected products If you have any affected products on stock, please quarantine them and send the products back to us. You will receive a replacement as soon as we have received your "Acknowledgement and Receipt Form" and know your current stock. Check any products or records of the affected LOTs used prior to receipt of this field safety notice if the tips have been intact after the surgery. If the tips have not been intact, ensure that any remaining parts are removed from the patient. There are no further specific patient monitoring actions necessary related to this Field Safety Notice. If after reviewing this notification you have any further questions or queries please discuss them with your pro med instruments sales representative Please complete the attachment "Acknowledgement and Receipt Form" and return by Fax or email (see under nr. 8. below) to pro med instruments GmbH to confirm receipt by the 10th of November 2017 at the latest."
Initiated
2017-11-02
Root cause
Under Investigation by firm
Terminated
2018-07-11
Distribution
Worldwide Distribution - US (nationwide) Internationally to Germany, Saudi Arabia, France, United Kingdom, Australia, Austria, Ireland, South Africa, Hungary, India, Spain, Italy, Japan, and Canada
Recall number
Z-0193-2018
Recall status
Terminated
CDRH record ID
159669
FDA record created
2017-12-05
Recall event number
78501

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗