Open original FDA record ↗Recall record
- FDA identifier
- Z-0193-2022
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA record creation date
- Nov 2, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The Aptio Automation Firmware for the ADVIA Centaur XP/XPT Interface Module may lead, in some cases, to an incorrect association of test results to sample ID. Incorrect association of test results could impact clinical interpretations and clinical decisions.
Source facts
- Action
- On October 6, 2021, the firm notified affected customers via an "Urgent Medical Device Correction" letter. Customers were informed of the product issue and provided instructions on actions to be taken by laboratories, which prevent the issue from happening. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative. A new version of the Firmware for the Aptio Automation Interface Module to the ADVIA Centaur XP/XPT will correct the issue. Your service provider will contact you to schedule the firmware upgrade.
- Initiated
- 2021-10-06
- Root cause
- Software design
- Distribution
- Worldwide distribution.
- Recall number
- Z-0193-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189788
- FDA record created
- 2021-11-02
- Recall event number
- 88865