Patients could possibly fall off the table if not properly secured when: 1) moving them onto the patient table, 2) moving them on the patient table or 3) removing them from the patient table.
Source facts
Action
The recalling firm issued a Customer Safety Advisory Notice dated 7/04/08 (including an Addendum to the Operators Manual) which alerts the operator of possible injury to patients if they are not properly secured to the table during examinations and describes the recommended fastening procedures required to properly secure the patient to the table.
For questions or additional information, contact Siemens Medical Solutions, Inc. at 610-448-4634.
Initiated
2008-08-22
Root cause
Labeling design
Terminated
2009-09-23
Distribution
Nationwide Distribution.
Recall number
Z-0194-2009
Recall status
Terminated
CDRH record ID
73744
FDA classified
2008-11-04
Recall event number
49529