Open original FDA record ↗Recall record
- FDA identifier
- Z-0195-2015
- Company in FDA record
- GE OEC Medical Systems, Inc
- FDA classification date
- Nov 7, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
GE Healthcare initiated a field correction because prior revision of components were potentially installed on certain OEC 9800, FlexiView 8800, OEC MiniView 6800 and OEC UroView 2800 systems.
Source facts
- Action
- On October 3. 2014. GE Healthcare Surgery began distributing an Urgent Medical Device Correction letter to customers to inform them of the issue and provide instructions for immediate mitigation of the issue. The letter was sent to 3 titles at each facility affected: 1.1 Facility Administrator. 2.1 Director/Manager of Radiology and Radiologists, 3.) Radiology Department. Firm will track all service requests for the field action to completion. These service requests will document the inspection and component replacement (if applicable) on the affected units. Exceptions to completion of the service request will be recorded. Records of field action completion will be retained per our processes and procedures. No product is being returned.
- Initiated
- 2014-10-03
- Root cause
- Component design/selection
- Terminated
- 2015-03-23
- Distribution
- Worldwide Distribution.
- Recall number
- Z-0195-2015
- Recall status
- Terminated
- CDRH record ID
- 130741
- FDA classified
- 2014-11-07
- Recall event number
- 69474