Open original FDA record ↗Recall record
- FDA identifier
- Z-0195-2023
- Company in FDA record
- Avanos Medical, Inc.
- FDA classification date
- Nov 8, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sterile extension sets were distributed without an expiration date
Source facts
- Action
- AVANOS sent an URGENT: FIELD CORRECTION on 09/15/2022 via overnight FedEx to customers. The letter explained the issue and requested the following: "WHAT SHOULD I DO IN RESPONSE TO THIS FIELD CORRECTION? Our records show that you and/or your facility have received one or more of the impacted products. Avanos requests that you take the following actions: " CHECK all storage and usage locations to determine if any impacted product remains within your possession and QUARANTINE all impacted product identified. " COMPLETE and RETURN the attached Acknowledgement Form (Attachment 1) to Avanos. o Via email to fieldactioncare@avanos.com , subject line FCA-2022-007 Hospital Acknowledgement Form o Avanos Customer Service will contact you regarding product return Please respond within five (5) business days of receipt of this letter. If you have questions or require further assistance, please contact Avanos via email at fieldactioncare@avanos.com." The firm sent updated letters the week of October 10, 2022. The letters titled, "URGENT: PRODUCT RECALL" provided updated product descriptions.
- Initiated
- 2022-09-14
- Root cause
- Incorrect or no expiration date
- Terminated
- 2023-08-31
- Distribution
- US: CA, CO, FL, MI, NM, OH, OR, TN, TX, UT, WI
- Recall number
- Z-0195-2023
- Recall status
- Terminated
- CDRH record ID
- 196309
- FDA classified
- 2022-11-08
- Recall event number
- 90964