Potential weld defect in the Overhead Tube Support assembly which connects to the telescoping arm. This could cause the tube assembly to fall, leading to possiible injury of the user or patient.
Source facts
Action
Analogic Corp notified accounts by email on 10/14/2005 and by letter on 10/17/2005. Accounts are requested to stop use and correction will be conducted by a service engineer.
Initiated
2005-10-14
Root cause
Other
Terminated
2015-04-14
Distribution
AZ, CA, FL, MA, OH, PA, TN, TX,
Recall number
Z-0196-06
Recall status
Terminated
CDRH record ID
42766
FDA classified
2005-11-24
Recall event number
33939