Skip to content

Public FDA record

Z-0196-2013: The touchscreen display (model 94267) designed for use with the XPREZZON bedside monitor (model 91393) is

Recall record Z-0196-2013 for Spacelabs Healthcare, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0196-2013
Company in FDA record
Spacelabs Healthcare, Llc
FDA classification date
Nov 5, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation.

Source facts

Action
Spacelabs Healthcare sent a URGENT - MEDICAL DEVICE CORRECTION letter dated October 19, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are advised to advise their staff of this potential failure and advise them to ensure the alarm lights are functional. The firm will contact the consignee at the earliest possible date to schedule a convenient time for the hospital to upgrade the affected product at no cost. Customers can call Technical Support at 1-800-522-7025 and select 2 for any questions about this recall.
Initiated
2012-10-05
Root cause
Vendor change control
Terminated
2013-03-25
Distribution
Worldwide Distribution--US (nationwide) including states of Alaska, California, Florida, Illinois, Michigan, South Carolina, and Texas., and the countries of Canada, Chile, France, Libya, Panama, Qatar, Saudi Arabia, Switzerland and United Kingdom
Recall number
Z-0196-2013
Recall status
Terminated
CDRH record ID
114078
FDA classified
2012-11-05
Recall event number
63506

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗