Open original FDA record ↗Recall record
- FDA identifier
- Z-0196-2013
- Company in FDA record
- Spacelabs Healthcare, Llc
- FDA classification date
- Nov 5, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation.
Source facts
- Action
- Spacelabs Healthcare sent a URGENT - MEDICAL DEVICE CORRECTION letter dated October 19, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are advised to advise their staff of this potential failure and advise them to ensure the alarm lights are functional. The firm will contact the consignee at the earliest possible date to schedule a convenient time for the hospital to upgrade the affected product at no cost. Customers can call Technical Support at 1-800-522-7025 and select 2 for any questions about this recall.
- Initiated
- 2012-10-05
- Root cause
- Vendor change control
- Terminated
- 2013-03-25
- Distribution
- Worldwide Distribution--US (nationwide) including states of Alaska, California, Florida, Illinois, Michigan, South Carolina, and Texas., and the countries of Canada, Chile, France, Libya, Panama, Qatar, Saudi Arabia, Switzerland and United Kingdom
- Recall number
- Z-0196-2013
- Recall status
- Terminated
- CDRH record ID
- 114078
- FDA classified
- 2012-11-05
- Recall event number
- 63506