Skip to content

Public FDA record

Z-0196-2019: Medacta International MiniMAX CEMENTLESS ANATOMICAL STEM, 12/14: for cementless use in total or partial h

Recall record Z-0196-2019 for Medacta Usa Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0196-2019
Company in FDA record
Medacta Usa Inc
FDA record creation date
Oct 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.

Source facts

Action
The firm initiated the recall by telephone to its distributors. The firm requested return of all recalled units.
Initiated
2017-11-15
Root cause
Package design/selection
Terminated
2019-03-29
Distribution
NC, ID, Belgium, Switzerland, Spain, Germany, Australia, France, Italy
Recall number
Z-0196-2019
Recall status
Terminated
CDRH record ID
167435
FDA record created
2018-10-19
Recall event number
81090

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗