Z-0196-2023: Disc Preparation instruments contained in the following products: CI-11050-08102515 Longer IdentiTi PO Tr

Recall record Z-0196-2023 for Alphatec Spine, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0196-2023
Company in FDA record
Alphatec Spine, Inc.
FDA classification date
Nov 8, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing instruments which included additional features and adequate consideration was not given to the need for a new Regulatory filing.

Source facts

Action
On 10/19/2022, the firm initially emailed its customer to inform them that Disc Preparation Instruments were developed, manufactured, and distributed that included an additional feature and the product did not go through the appropriate regulatory Pre-Market filing. Customer are instructed to: 1) Promptly review their inventory to determine if any of the affected product listed above is within their possession. 2) Abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3) Fill out the last page of this letter to confirm that they have read this notification and have taken all necessary actions as described in this notification. 4) Return page 4 (signed) of this letter using one of the methods below: -Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Mihir Naik -Email to: QMSCompliance@atecspine.com For questions, contact 1-800-922-1356
Initiated
2022-09-14
Root cause
No Marketing Application
Distribution
U.S.: CA O.U.S.: N/A
Recall number
Z-0196-2023
Recall status
Open, Classified
CDRH record ID
196239
FDA classified
2022-11-08
Recall event number
90940

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗