The STERRAD 100 and the STERRAD 100S vaporizer plates are prone to displacement allowing micro-droplets of H2O2 to contact the load. If the vaporizer plate is dislodged, there is an increased chance for operators to have contact with H2O2 resulting in transient burns during the removal of the load.
Source facts
Action
The recall strategy consists of supplying each customer of the STERRAD 100 and the STERRAD 1OOS with: a notification letter of the field action (mailed 09/22/05 ), a pack containing two vaporizer plates of the newly released design, and a set of instructions for self-installation.
Initiated
2005-09-16
Root cause
Other
Terminated
2011-11-21
Distribution
Nationwide and Canada, Germany, & Japan.
Note: A more detail list of international consignees
pending.
Recall number
Z-0197-06
Recall status
Terminated
CDRH record ID
41983
FDA classified
2005-11-24
Recall event number
33513