Open original FDA record ↗Recall record
- FDA identifier
- Z-0197-2007
- Company in FDA record
- Beckman Coulter Inc
- FDA classification date
- Nov 21, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Mis-identification-If a panel or protocol is added to an existing worklist but the tube location is not specified, the CXP Acquisition software will run the last specified tube through the remaining protocols and will generate results and printouts of these runs leading to a Mis-Identification condition.
Source facts
- Action
- Customer letter -- A Product Corrective Action (PCA) letter was sent the week of July 24, 2006 to all accounts that have FC500 with CXP Software 2.0 and 2.1 to inform them that a potential sample misidentification may occur when running a Work list with missing tube Location(s) in Automatic MCL mode. The letter outlines actions to be taken to avoid mis-identification of patient results.
- Initiated
- 2006-07-26
- Root cause
- Other
- Terminated
- 2012-02-29
- Distribution
- Nationwide and Canada
- Recall number
- Z-0197-2007
- Recall status
- Terminated
- CDRH record ID
- 48584
- FDA classified
- 2006-11-21
- Recall event number
- 36397