A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.
Source facts
Action
The recalling firm will notify their consignees via letter delivered by traceable courier service. The consignee will be asked to complete and return a form to the recalling firm stating if they have affected product or not.
Initiated
2014-10-24
Root cause
Nonconforming Material/Component
Terminated
2015-07-21
Distribution
Nationwide Distribution including the states of TX, CA, OH, MS, VA, IL, TN, GA, CA, FL, and AZ.
Recall number
Z-0197-2015
Recall status
Terminated
CDRH record ID
131021
FDA classified
2014-11-08
Recall event number
69596