Z-0197-2019: TRINIAS Digital Angiographic System intended to be used for cardiac angiography, neurovascular angiograph

Recall record Z-0197-2019 for Shimadzu Medical Systems Usa Com. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0197-2019
Company in FDA record
Shimadzu Medical Systems Usa Com
FDA classification date
Oct 19, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a software issue the C-arm may unexpectedly moved when being moved to the park position and during specific operations. There is a possibility the C-arm will come into contact with a patient or healthcare personnel.

Source facts

Action
On June 15, 2018, the firm sent an URGENT Voluntary Medical Device Recall Notice to its customer via certified mail informing the customer that the Trinias C-arm could unexpectedly move while being moved to the park position and/or when performing specific operations. An investigation has shown that a software defect in the MH300, which is a floormount C-arm tube support of the Trinias biplane system, caused this event under specific conditions described as below. Neither the Trinias single plane system nor the Trinias dual plane system have the software defect. If the incident were to recur, there is a possibility the C-arm will come into contact with a patient or healthcare personnel. The specific conditions are as follows: (1) The table top is positioned at head side so that the C-arm safety interlock is active. (2) If operator performs a parking operation and presses a direct memory switch at the same time, the the software runs abnormally, and the C-arm does not movebecause the safety interlock is active (1). (3) If the operator then moves the table top to foot side, the C-arm safety interlock is cancelled. (4) While the C-arm safety interlock is cancelled due to the actions performed in (3), the C-arm starts moving without the operators control. Therefore, the firm will implement a voluntary recall/field corrective action to revise the software, thereby fixing the problem. This Field Safety Notice is intended to inform the customer about: - the description of the issue - the actions planned by Shimadzu to correct the issue - the actions that should be taken by the customer How to identify the affected products: The Shimadzu Digital Angiographic System Trinias; Serial Number 41E58C977001 Please contact the local Shimadzu representative or contact Shimadzu National Technical Support. Corrective action plan by Shimadzu: After the software defect has been corrected and the correction safety and efficacy have been verified through our verifi
Initiated
2018-06-15
Root cause
Software design
Terminated
2021-04-28
Distribution
Domestic: LA
Recall number
Z-0197-2019
Recall status
Terminated
CDRH record ID
167416
FDA classified
2018-10-19
Recall event number
81080

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗